An inversion recovery sequence begins with a 180 degree pulse that flips longitudinal magnetization to the negative z axis. During the delay TI, T1 relaxation brings it back toward equilibrium, and a 90 degree pulse then reads out the current value. Tissue signal passes through zero at the null point, TI = T1 ln2, which is used to suppress fat (STIR) or fluid (FLAIR).
Intuition
Picture the magnetization as a vector that is flipped upside down and then slowly climbs back up. Because it starts at minus and ends at plus , it must cross zero on the way. The time of that crossing depends only on . Tissues with short recover quickly and cross zero early, while tissues with long cross later. If you fire the 90 degree readout pulse exactly when one tissue crosses zero, that tissue contributes no signal, and the other tissues, which are at different points on their recovery curves, still do. The sim lets you drag TI and watch each tissue's curve and the resulting image brightness.
Key equations
Longitudinal magnetization at the time of the 90 degree readout, assuming full recovery between repetitions (long TR).
Inversion time at which Mz crosses zero and that tissue gives no signal.
General form when TR is finite, so recovery is incomplete.
Derivation
- After the 180 degree pulse, .
- recovery obeys , giving .
- Substitute : .
- Set Mz = 0: 2e^{-TI/T1} = 1, so e^{-TI/T1} = 1/2.
- Take logs: TI/T1 = ln2, so TI_null = T1 ln2.
- With finite TR, the starting value is not but approximately, which adds the term and shifts the null to a shorter TI.
Worked example
Fat at 1.5 T has T1 about 260 ms. Find the TI that nulls fat (STIR), assuming long TR. Then find the null for CSF with T1 about 4000 ms (FLAIR).
- Fat: TI = 0.693 x 260 ms = 180 ms.
- CSF: TI = 0.693 x 4000 ms = 2772 ms, about 2.8 s.
- Check fat at TI = 180 ms: 1 - 2e^{-0.692} = 1 - 2(0.5) = 0, as expected.
- In practice, FLAIR uses a TI near 2200 ms because finite TR shortens the null.
Answer Fat null is about 180 ms (STIR); CSF null is about 2.8 s with long TR, and roughly 2.2 s in practice with finite TR (FLAIR).
Common mistakes
- Using TI = T1 instead of T1 ln2 for the null.
- Forgetting that the finite-TR formula shifts the null to shorter TI.
- Thinking the signal is negative: magnitude images show |Mz|, so tissue just before and after the null look equally dark near it, not negative.
- Confusing TI (inversion to readout) with TR (between repetitions) or TE.
- Assuming STIR nulls fat by chemical shift; it relies on T1, so other short-T1 tissues, such as gadolinium-enhanced ones, can also be nulled.