Content
Summary
In conventional spin-warp MRI, k-space is filled one horizontal line per repetition time (TR). Each line has a fixed phase-encode step set by a gradient pulse, and the readout gradient sweeps along kx while the signal is sampled. Total scan time is therefore the product of TR, the number of phase-encode lines, and the number of averages.
Intuition
Picture k-space as a grid, a map of spatial frequencies. The readout gradient moves you left to right along a row, like a pen drawing a line while the ADC samples. The phase-encode gradient decides which row you are on, by moving you up or down before the readout starts. After the readout the sequence must wait for the next TR, then it takes another step in ky and draws the next line. Rows near the center (ky near 0) carry most signal and contrast. Outer rows carry fine detail and edges. The slow part is that only one row can be acquired per excitation, so image height in pixels sets the number of TRs.
Key equations
Position along kx grows with the accumulated readout gradient area.
Phase-encode gradient amplitude (or duration) selects the ky row.
Spacing between phase-encode lines sets the field of view in y.
Scan time for a 2D spin-warp acquisition with N_pe lines and NEX averages.
Derivation
- Spins in a gradient precess at a frequency that depends on position: .
- Phase accumulated is , which defines , so signal .
- Applying Gy for a fixed time Tpe before readout leaves each y position with phase 2π ky y, with ky set by Gy amplitude.
- Stepping Gy in equal increments from -Gmax to +Gmax gives Npe equally spaced ky values with spacing Δky.
- The Fourier relation says sample spacing Δk gives FOV=1/Δk, and the maximum extent kmax gives resolution Δy=1/(2kmax).
- Since one excitation yields one readout, one ky line is acquired per TR, so total time is Npe×TR×NEX.
Worked example
A 2D spin-echo scan has a 256 x 256 matrix, TR = 500 ms, and NEX = 2. Phase encoding is along y. Find the scan time and the ky spacing for FOV_y = 240 mm.
- Number of phase-encode lines Npe = 256.
- Scan time = 256 x 0.5 s x 2 = 256 s.
- Convert to minutes: 256/60 ≈ 4.27 min.
- Δky = 1/FOV = 1/0.24 m = 4.17 cycles/m (0.0417 cycles/cm).
- Maximum ky = (Npe/2)Δky = 128 x 4.17 ≈ 533 cycles/m, giving pixel size 240/256 = 0.94 mm.
Answer About 256 s (4 min 16 s); Δky ≈ 4.17 cycles/m.
Common mistakes
- Thinking each TR fills the whole of k-space rather than a single line.
- Confusing the frequency-encode direction (sampled continuously during readout) with phase-encode (stepped between TRs).
- Believing that more readout samples along kx lengthens scan time; it does not, but more ky lines does.
- Assuming the outer lines carry the most signal; the center of k-space dominates contrast and signal.
- Forgetting that the FOV is set by line spacing, not by the maximum ky extent.