Content
Summary
A gradient echo sequence uses a low flip angle excitation and a reversal of a magnetic field gradient to refocus signal, without a 180 degree pulse. Because static field inhomogeneities are not refocused, signal decays with T2* rather than T2, making the sequence fast but sensitive to susceptibility effects.
Intuition
After excitation, spins in the transverse plane precess at slightly different frequencies. Some difference comes from random spin-spin interactions (true T2), and some from fixed field inhomogeneities, such as air-tissue interfaces or iron in blood products. A gradient pulse deliberately dephases spins; reversing it rewinds only that deliberate dephasing, producing an echo. Fixed inhomogeneity dephasing keeps accumulating, so the echo peak is smaller the longer TE is. A small flip angle leaves much longitudinal magnetization intact, allowing short TR and fast imaging. The simulation lets students watch a fan of spin vectors dephase, see the gradient flip bring them into alignment, and compare the echo envelope with T2* decay.
Key equations
T2* relaxation rate is the true T2 rate plus a contribution from field inhomogeneity (ΔB0 is the spread in field; the term is expressed in rate units).
Gradient echo signal amplitude at echo time TE decays with T2*.
Ernst angle: flip angle that maximizes steady-state signal for given TR and T1.
Spoiled gradient echo steady-state signal.
Derivation
- After excitation, each spin has phase φ(t)=ω t, where ω includes the Larmor frequency offset from inhomogeneity and any applied gradient.
- Without a gradient, offsets from ΔB0 produce a spread of phases, so net transverse magnetization falls faster than by T2 alone: 1/T2* = 1/T2 + 1/T2'.
- Apply a negative readout gradient lobe: spins dephase proportional to position, φ=-γGx t.
- Reverse the gradient to positive: the gradient-induced phase accumulates back, reaching zero at time TE when the areas of the two lobes are equal.
- At that moment the gradient dephasing is cancelled, giving an echo, but the dephasing from ΔB0 has not been reversed, since no 180 degree pulse acted.
- Therefore the echo peak amplitude follows .
- Using steady-state Bloch solutions with spoiled transverse magnetization gives the signal equation, maximized at the Ernst angle.
Worked example
A gradient echo scan has T2 = 80 ms and T2' = 40 ms (field inhomogeneity contribution). Find T2*, and the signal fraction remaining at TE = 20 ms compared with TE = 5 ms.
- Rate: 1/T2* = 1/80 + 1/40 = 0.0125 + 0.025 = 0.0375 per ms.
- T2* = 1/0.0375 = 26.7 ms.
- Signal at TE = 5 ms: exp(-5/26.7)=exp(-0.1875)=0.829.
- Signal at TE = 20 ms: exp(-20/26.7)=exp(-0.75)=0.472.
- Ratio = 0.472/0.829 = 0.57.
Answer ms; the signal at TE 20 ms is about 47% of , roughly 57% of that at TE 5 ms.
Common mistakes
- Saying the gradient reversal refocuses all dephasing; it refocuses only the gradient-induced part.
- Assuming T2* is longer than T2; it is always shorter or equal.
- Using a 90 degree flip angle with very short TR, which saturates the signal.
- Forgetting that susceptibility artifacts (signal voids near air or metal) grow with TE.
- Confusing gradient echo with spin echo, where a 180 degree pulse does refocus static inhomogeneity.