Content
Summary
Echo-planar imaging (EPI) fills all of k-space after a single RF excitation by rapidly oscillating the readout gradient and adding small phase-encode 'blips' between readout lines. The resulting zigzag trajectory acquires a full image in tens of milliseconds, but it is sensitive to off-resonance and T2* decay, producing distortion and blurring along the phase-encode direction.
Intuition
Imagine walking across a field in rows. In conventional imaging each row needs its own excitation and a wait of TR. In EPI you excite once, then walk a row left to right, take one small step up (the blip), walk right to left, step up again, and so on, all while the signal decays with T2*. The readout gradient alternates polarity, so odd lines are traversed in reverse. The blip is a short phase-encode gradient lobe whose area sets the k-space step in ky. Because the time between adjacent ky lines is the echo spacing (about 0.5 ms) while the time between adjacent kx points is only microseconds, any frequency error accumulates mostly along ky. A field offset of a few tens of Hz therefore shifts pixels by a noticeable number of voxels in the phase direction, while the frequency direction is almost unaffected.
Key equations
k-space step along readout set by gradient amplitude and dwell time
Each blip area gives one ky step, fixing the phase-direction field of view
Readout train length equals number of lines times echo spacing
Off-resonance Δf shifts pixels along phase direction by Δf times total readout time, times FOV
Phase-direction bandwidth per pixel is very low, explaining distortion
Derivation
- Position in k-space is the time integral of gradient area: k(t) = (γ/2π)∫G dt.
- After excitation, a pre-phasing lobe moves the start point to (−kx,max, −ky,max).
- A readout gradient of +Gx sweeps kx across its range while samples are acquired.
- At the line end, a short blip Gy moves ky by Δky = 1/FOVy, with no sampling during the blip (or sampling on ramps).
- The readout gradient reverses to −Gx, sweeping kx back; this is repeated Ny times, forming a zigzag.
- An off-resonance Δf adds extra phase 2πΔf t. Between adjacent ky lines this is 2πΔf·ESP, equivalent to an apparent Δky shift.
- Total phase-direction shift in pixels = Δf · Ny · ESP, since pixel bandwidth is 1/(Ny·ESP).
- Reversed readout lines also cause odd-even misregistration, producing Nyquist (N/2) ghosts when gradient delays or eddy currents are present.
Worked example
A 64×64 EPI acquisition has echo spacing 0.5 ms and FOV 240 mm. Find the total readout time, phase-direction bandwidth per pixel, and the shift in mm for a 100 Hz off-resonance region (e.g., near a sinus).
- Readout time T = 64 × 0.5 ms = 32 ms.
- Bandwidth per pixel = 1/T = 1/0.032 s ≈ 31.25 Hz.
- Shift in pixels = 100 / 31.25 = 3.2 pixels.
- Pixel size = 240/64 = 3.75 mm.
- Shift = 3.2 × 3.75 ≈ 12 mm.
Answer T ≈ 32 ms, ≈31 Hz per pixel, and a displacement of about 3.2 pixels ≈ 12 mm along the phase-encode direction.
Common mistakes
- Thinking the blip moves along kx; it moves only along ky.
- Assuming distortion occurs along the readout direction; it is mainly along phase encode.
- Forgetting that alternate lines are traversed in reverse and must be flipped before Fourier transform.
- Confusing T2* decay with T2; EPI signal weights follow T2* across the train.
- Believing that a faster readout gradient alone removes distortion; reducing it requires shorter ESP, parallel imaging, or fewer lines.
- Ignoring that the blip area, not its duration, sets the ky step.