Content
Summary
A spiral k-space trajectory samples the spatial-frequency plane along a curve that winds outward from the center, driven by sinusoidal, amplitude-ramped gradient waveforms. It covers k-space efficiently with smooth gradients, enabling fast imaging, but is sensitive to off-resonance blurring and requires gridding for reconstruction.
Intuition
In MRI the position in k-space is the time integral of the gradient waveform: . If both x and y gradients oscillate sinusoidally with slowly growing amplitude, with a 90 degree phase shift, the k-space point traces a spiral. Unlike echo-planar imaging, there are no sharp turnarounds, so the gradient hardware is used gently and continuously. Starting at the center, where most signal and contrast energy lies, each turn adds a ring of higher spatial frequencies. The limits are the maximum gradient amplitude and the maximum slew rate. Early in the spiral the slew rate limits how fast k can grow; later the amplitude limit takes over. The price of efficiency is that field inhomogeneity makes phase accumulate during a long readout, blurring the image instead of distorting it geometrically.
Key equations
k-space position is the integrated gradient waveform.
Radius grows linearly with angle; N is the number of interleaves-adjusted turns, with spacing 1/FOV per turn for single shot.
Complex notation for the spiral path.
Gradients follow from the trajectory and must respect hardware limits.
Radial spacing between turns that avoids aliasing within the field of view.
Derivation
- Start with .
- Choose an Archimedean spiral: , so radius increases by per turn.
- Require turn spacing equal to 1/FOV for each interleave: 2piA = 1/(FOV n_int), giving A = 1/(2pi FOV n_int).
- Differentiate to get the gradient: is proportional to .
- For constant angular rate, gradient magnitude grows with and slew rate is limited early, so is chosen to satisfy the slew limit.
- When reaches , solve for so that is constant, which is the amplitude-limited regime.
- The readout ends when k reaches kmax = 1/(2 Δx), set by the desired resolution.
Worked example
A single-shot spiral has FOV = 24 cm and resolution 2 mm (Δx = 0.2 cm). Find the number of turns needed, assuming Archimedean spacing of 1/FOV.
- kmax = 1/(2Δx) = 1/(0.4 cm) = 2.5 cm^-1.
- Turn spacing Δk = 1/FOV = 1/24 = 0.0417 cm^-1.
- Number of turns = kmax/Δk = 2.5 × 24 = 60.
- With 4 interleaves each spiral needs only 60/4 = 15 turns.
Answer About 60 turns for a single shot, or about 15 turns per interleave with 4 interleaves.
Common mistakes
- Thinking spiral gradients are zero at the start; they ramp up from zero, but k starts at the center.
- Sampling data uniformly in k-space and using a plain FFT; non-Cartesian data need gridding with density compensation.
- Ignoring slew-rate limits and designing only for gradient amplitude.
- Assuming off-resonance gives geometric distortion as in EPI; in spiral it gives blurring.
- Using turn spacing larger than 1/FOV, which causes aliasing artifacts.