Summary
The CPMG sequence applies a 90 degree pulse followed by a train of 180 degree refocusing pulses, producing a series of spin echoes whose peak amplitudes decay with the true transverse relaxation time T2. Fitting this decay gives T2 while the repeated refocusing cancels static field inhomogeneity and, with short echo spacing, reduces diffusion effects.
Intuition
After a 90 degree pulse, spins precess in the transverse plane at slightly different frequencies because of field inhomogeneity and chemical shift, so they fan out and the signal vanishes quickly (T2*). A 180 degree pulse flips the fan, so the fast spins that were ahead are now behind and catch up. At time 2tau after the excitation the spins realign and an echo forms. Reversible dephasing is undone, but irreversible dephasing from random molecular fluctuations (true T2 processes) cannot be undone. Each echo is therefore a little smaller than the last. CPMG applies the refocusing pulses with a 90 degree phase shift relative to the excitation pulse, so that pulse flip-angle errors cancel on even echoes instead of accumulating. Plotting echo peak heights against time gives a clean exponential decay.
Key equations
Echo amplitude at time t_n decays with the true T2.
The nth echo occurs at n times the echo spacing TE, which is twice the pulse spacing tau.
Taking logs gives a straight line with slope -1/T2.
Observed decay combines intrinsic relaxation and inhomogeneity; the echo removes the second term.
Derivation
- Right after the pulse, magnetization lies in the transverse plane with magnitude .
- During time , a spin with offset frequency accumulates phase .
- A degree pulse about an axis in the plane inverts this phase to minus times .
- During the next the spin gains the same phase again, so the net phase is zero at regardless of .
- Irreversible random-field processes give phases that differ in each interval, so they do not cancel; they cause amplitude loss at rate 1/T2.
- Repeating 180 pulses every 2tau gives echoes at 2tau, 4tau, 6tau, and so on, each multiplied by exp(-TE/T2) relative to the previous one.
- Therefore the th` echo is , and fitting the peaks yields .
Worked example
A CPMG train with TE = 20 ms gives echo amplitudes 80.0 at echo 1 and 29.4 at echo 5. Estimate T2 and the amplitude at echo 10.
- Time of echo 1 is 20 ms; time of echo 5 is 100 ms; the interval is 80 ms.
- Ratio of amplitudes: 29.4/80.0 = 0.3675.
- ln(0.3675) = -1.001.
- So 80 ms / T2 = 1.001, giving T2 = about 80 ms.
- Echo 10 is at 200 ms, 100 ms after echo 5: 29.4 × exp(-100/80) = 29.4 × 0.2865 = 8.4.
Answer T2 is about 80 ms, and the echo 10 amplitude is about 8.4 (arbitrary units).
Common mistakes
- Confusing T2 with T2*: a single free induction decay gives T2*, not T2.
- Using a TE that is too long, so diffusion in gradients shortens the apparent T2.
- Forgetting that the first echo is at TE, not at time zero, when fitting.
- Ignoring pulse imperfections or using the wrong pulse phase, causing odd-even echo amplitude oscillation.
- Fitting a single exponential to a tissue with several T2 components, such as myelin water and free water.