Content
Summary
Balanced SSFP (bSSFP, TrueFISP/FIESTA/bFFE) uses a fast train of RF pulses with fully refocused gradients on all axes, so transverse magnetisation is preserved from one TR to the next. The steady-state signal depends on / and on the off-resonance phase accrued per TR, producing characteristic dark bands where that phase equals odd multiples of .
Intuition
Imagine spins being kicked every by a pulse of flip angle , alternating in sign (0, 180 phase cycling) or equivalently viewed in a frame that handles the alternation. Between pulses, spins relax partly and precess by an angle , where is the off-resonance. Because gradients are balanced, no extra dephasing is added, so the magnetisation left over from one cycle becomes the starting point for the next. After a few times the system settles to a fixed point. Signal is high and depends on the ratio , giving bright fluid and blood. But the fixed point depends on : at on-resonance the signal is at a maximum, and at () the transverse components cancel destructively, giving a null: a banding artefact. Shortening widens the bands apart, pushing them away from the region of interest.
Key equations
Off-resonance phase accrued per repetition; banding repeats every 1/TR in Hz.
On-resonance bSSFP steady-state transverse signal (TR << T1,T2).
Flip angle that maximises signal; gives .
Off-resonance frequency at the centre of the first dark band ().
Derivation
- Per TR, magnetisation undergoes: RF rotation by about , then free precession by about with transverse decay , and longitudinal recovery with .
- Write the cycle as , where and after recovery.
- The steady state satisfies M_ss = (I - A)^(-1) B.
- Solving gives a transverse signal with numerator proportional to and a denominator that is a function of , , and .
- For on-resonance () and , this simplifies to the formula shown.
- Maximising over alpha gives the optimal flip angle; the signal then scales with , independent of absolute and values.
- The denominator depends on ; at the transverse contributions from successive pulses subtract, creating a signal minimum, which is the band.
Worked example
A bSSFP sequence uses TR = 4 ms. Where do the dark bands fall, and what is the optimal flip angle for tissue with T1 = 1200 ms and T2 = 40 ms?
- Band centres are at , so .
- Bands repeat every 1/TR = 250 Hz, so at +/-125 Hz, +/-375 Hz, and so on.
- T1/T2 = 30. cos(alpha_opt) = (30-1)/(30+1) = 29/31 = 0.935.
- alpha_opt = arccos(0.935) ≈ 20.6 degrees.
- Peak signal is .
Answer Dark bands at about +/-125 Hz (repeating every 250 Hz); optimal flip angle about 21 degrees for this tissue. For fluids with T2/T1 near 1 the optimum is nearer 90 degrees.
Common mistakes
- Believing bSSFP contrast is T1-weighted; it is mainly T2/T1 weighted.
- Thinking banding is caused by T2* dephasing within a voxel; it arises from off-resonance phase accumulation across a TR.
- Lengthening TR to reduce banding; that brings bands closer together, so you should shorten TR.
- Forgetting that the 0/180 phase cycling shifts the band pattern by half a period, so the signal at on-resonance is a maximum.
- Using the T2/T1 formula when TR is not short compared with T1 and T2.