A spoiled gradient echo (SPGR/FLASH) sequence repeats a low flip angle pulse every TR and destroys residual transverse magnetization each cycle, so only longitudinal recovery carries forward. The signal reaches a steady state that depends on T1, TR and flip angle, and is maximal at the Ernst angle, cos(alpha_E) = exp(-TR/T1).
Intuition
Each RF pulse tips some longitudinal magnetization into the transverse plane, where it gives signal; the rest stays along z. If TR is short compared with T1, z magnetization has little time to recover, so a large flip angle quickly saturates the spins and the signal collapses after a few pulses. A small flip angle uses only a fraction of Mz each time, leaving more for the next pulse, but yields less signal per pulse. The best compromise is the Ernst angle: small enough to preserve Mz, large enough to give transverse signal. Spoiling (gradient dephasing plus RF phase cycling) ensures no transverse coherence builds up, so the simple T1-only model holds. Short TR relative to T1 means a small Ernst angle; as TR approaches several T1, the Ernst angle approaches 90 degrees.
Key equations
Steady-state spoiled gradient echo signal
Ernst angle that maximizes the signal
Effect of pulse and recovery over one TR
Small-TR approximation, in radians
Derivation
- Assume perfect spoiling: transverse magnetization is zero just before each pulse.
- Just before a pulse, longitudinal magnetization is . The pulse leaves and creates transverse magnetization .
- During TR, recovers: , with .
- At steady state , so .
- Signal is proportional to decay at TE, giving the signal equation.
- Set : , so .
Worked example
Brain tissue with T1 = 1000 ms is imaged with spoiled GRE at TR = 20 ms. Find the Ernst angle and the relative signal at the Ernst angle versus at 90 degrees.
- E1 = exp(-20/1000) = 0.9802.
- cos(alpha_E) = 0.9802, so alpha_E = arccos(0.9802) ≈ 11.4 degrees.
- Signal factor at Ernst: (times ).
- At 90 degrees: (times ).
- Ratio ≈ 0.0999/0.0198 ≈ 5.
Answer Ernst angle ≈ 11°; its signal is about five times that at 90 degrees (≈0.10 M0 vs 0.02 M0).
Common mistakes
- Using the Ernst angle formula without spoiling; in unspoiled steady state (e.g. SSFP) T2 matters and the formula fails.
- Thinking the Ernst angle maximizes contrast; it maximizes signal of one T1, while T1 weighting needs a larger angle.
- Forgetting that the Ernst angle depends on tissue T1, so it is optimal for only one tissue.
- Mixing degrees and radians in the small-angle approximation.
- Ignoring imperfect spoiling or B1 inhomogeneity, which make the actual flip angle differ from nominal.