Content
Summary
In MRI, a slice is chosen by applying a linear gradient along one axis so that resonance frequency varies with position, while a shaped radiofrequency (RF) pulse with a limited frequency band excites only the spins whose frequencies fall in that band. A sinc-shaped pulse has a nearly rectangular frequency spectrum, so it excites a sharp, flat slice. Slice thickness depends on RF bandwidth and gradient strength; slice position depends on the RF center frequency.
Intuition
Think of the patient as a long row of tuning forks, all identical at rest. The slice gradient makes each fork slightly different in pitch according to where it sits along the axis. An RF pulse is like a sound containing only a narrow range of pitches. Only forks tuned within that range ring, so only one slab of tissue responds. The Fourier transform links the pulse shape in time to its frequency content: a sinc in time gives a rectangle in frequency, which gives a rectangle of excited positions. Making the pulse more narrowband, or the gradient weaker, gives a thicker slice. Moving the carrier frequency slides the slice along the gradient axis.
Key equations
Larmor frequency varies linearly with position z in the presence of the slice gradient G_z.
Slice thickness equals RF bandwidth divided by times gradient strength (, about 42.58 MHz/T for protons).
Slice center is set by the offset of the RF carrier frequency from the central Larmor frequency.
Small-flip-angle excitation profile is approximately the Fourier transform of the RF envelope.
Derivation
- With gradient on, spins at position precess at .
- In the small-flip-angle approximation, the transverse magnetization profile is approximately the Fourier transform of B1(t).
- A sinc envelope with main-lobe-defined bandwidth Δf has a rectangular spectrum of width Δf centered on the carrier frequency fc.
- Spins are excited where f(z) lies within fc ± Δf/2.
- Solving gives center and thickness .
- In practice the sinc is truncated and apodized (e.g., Hamming window), and a rephasing gradient lobe of opposite sign, half the area, follows to refocus phase across the slice.
Worked example
A 1 ms sinc pulse has a bandwidth of 2 kHz. The slice gradient is 10 mT/m. What is the slice thickness for protons, and what carrier offset moves the slice 3 cm from isocenter?
- .
- Gradient in frequency per distance: 42.58 kHz/mT × 10 mT/m = 425.8 kHz/m = 0.4258 kHz/mm.
- Thickness = 2 kHz / 0.4258 kHz/mm ≈ 4.7 mm.
- Offset for 30 mm shift = 0.4258 kHz/mm × 30 mm ≈ 12.8 kHz.
Answer Thickness about 4.7 mm; carrier offset about +12.8 kHz (sign depends on gradient polarity).
Common mistakes
- Thinking the RF pulse itself is spatially selective; it is the gradient that maps frequency to position.
- Forgetting that a longer pulse has narrower bandwidth, giving a thicker slice at fixed gradient.
- Confusing with and mixing rad/s with Hz.
- Omitting the rephasing gradient lobe, which causes signal loss from dephasing across the slice.
- Assuming a real truncated sinc gives perfectly sharp edges; the profile has ripples and transition bands.