In a pulse-acquire NMR experiment, a short radiofrequency pulse tips the net magnetization from the z axis into the transverse plane. The magnetization then precesses and decays, inducing an oscillating signal in the coil called the free induction decay (FID). A Fourier transform converts the FID into a spectrum with peaks at the resonance frequencies.
Intuition
At equilibrium in a strong field , spins give a small net magnetization along . This cannot be detected directly because the coil senses only changing flux in the transverse plane. A 90 degree pulse at the Larmor frequency rotates onto the plane. There it spins like a tiny bar magnet, inducing a voltage in the coil. Over time the spins lose phase coherence ( or relaxation) and the signal shrinks, while slowly regrows along (). The recorded signal is a decaying cosine, or a sum of several. Fourier transform asks which frequencies are present and how much of each. A single decaying oscillation gives one peak, its width set by the decay rate. Faster decay means a broader line. Several chemical environments give several peaks at different offsets, measured in ppm.
Key equations
Larmor frequency: precession rate of spins in field , with gyromagnetic ratio .
Flip angle from pulse amplitude and duration ; a 90 degree pulse has .
FID in the rotating frame, offset from the reference frequency, decaying with T2*.
Full width at half height of the Lorentzian line after Fourier transform.
Chemical shift in ppm, independent of field strength.
Derivation
- Start with along at equilibrium.
- Apply a field rotating at the Larmor frequency along in the rotating frame; precesses about at rate .
- After time the angle is ; choose it to equal 90 degrees so lies along .
- After the pulse, in the rotating frame precesses at the offset and the transverse part decays: .
- The coil detects Mxy, giving the FID S(t).
- Fourier transform: the integral of S(t)e^{-iωt}dt gives a Lorentzian centred at Δω with half-width 1/(πT2*).
- Peak area is proportional to the number of nuclei contributing.
Worked example
A proton FID, on a 400 MHz spectrometer, oscillates at 1200 Hz relative to the reference and decays with T2* = 0.25 s. Find the chemical shift and linewidth.
- Chemical shift = 1200 Hz / 400 MHz = 1200/400,000,000 = 3.0e-6, i.e. 3.0 ppm.
- Linewidth = 1/(π × 0.25 s) = 1/0.785 = 1.27 Hz.
- In ppm the width is 1.27/400 ≈ 0.003 ppm.
Answer Peak at 3.0 ppm with a width of about 1.3 Hz.
Common mistakes
- Thinking the signal is measured along z; only transverse magnetization is detected.
- Confusing T2 with T2*: field inhomogeneity makes the observed decay faster than true T2.
- Believing a 180 degree pulse gives maximum signal; it gives none, while 90 gives the maximum.
- Saying a longer FID gives a broader line; slow decay gives a narrower line.
- Forgetting that ppm is field-independent while Hz is not.