Content
Summary
The Ernst angle is the flip angle that maximizes steady-state signal in a spoiled gradient-echo or repeated-pulse experiment when the repetition time TR is short compared with T1. It depends only on the ratio TR/T1 and gives the best compromise between how much magnetization is tipped into the transverse plane and how much longitudinal magnetization has recovered between pulses.
Intuition
If you use a 90 degree pulse with a short TR, you wipe out the longitudinal magnetization each time, and it has little time to regrow, so later signals become small. If you use a tiny flip angle, the magnetization stays near equilibrium but you only tip a small fraction into the transverse plane, so the signal is small too. Somewhere in between is a best flip angle. The shorter TR is relative to T1, the smaller this optimal angle becomes. For TR much longer than T1, the optimum approaches 90 degrees.
Key equations
Fraction of longitudinal magnetization that remains after one repetition time as it recovers toward equilibrium.
Steady-state spoiled gradient-echo signal ( weighting ignored) as a function of flip angle .
Ernst angle condition that maximizes the steady-state signal.
Derivation
- Before a pulse, the steady-state longitudinal magnetization is . After the pulse it is , and the transverse magnetization is .
- During TR, with complete spoiling of the transverse magnetization, recovers: .
- Solve for : .
- Signal is proportional to the transverse magnetization: .
- Differentiate with respect to and set . The numerator of the derivative is .
- Setting this to zero gives cos(alpha_E) = E1 = exp(-TR/T1).
Worked example
Tissue has . A spoiled gradient-echo sequence uses . Find the Ernst angle and the relative signal compared with .
- Compute E1 = exp(-20/1000) = exp(-0.02) = 0.9802.
- The Ernst angle satisfies , so degrees.
- The signal is .
- Numerator: 0.0198 × 0.1977 = 0.003915.
- Denominator: 1 - 0.9608 = 0.0392.
- , about 10 percent of M_0.
Answer The Ernst angle is about 11 degrees, giving a steady-state signal of about 0.10 .
Common mistakes
- Assuming 90 degrees always gives the most signal; this is true only when TR is much longer than T1.
- Applying the formula without spoiling, or to balanced SSFP sequences, where the optimum depends on T2/T1 as well.
- Using the wrong tissue T1 or mixing units for TR and T1.
- Forgetting that maximizing signal from one tissue does not maximize contrast between tissues.
- Ignoring T2* and TE decay, which scale the signal but do not change the Ernst angle.