Content
Summary
Diffusion-weighted MRI makes the signal sensitive to random water motion by applying a pair of strong gradient pulses. The b-value combines gradient strength, duration and spacing into a single number that sets how strongly diffusion attenuates the signal. Higher b-values give more diffusion contrast but less signal.
Intuition
Imagine tagging water molecules with a phase 'label' that depends on position along a gradient. A second, opposite gradient then undoes the label. If molecules stayed still, the label would be perfectly undone and the signal would be unchanged. But molecules that wandered to a new position between the two pulses are unlabeled incorrectly, so their phases are scrambled and they contribute less to the net signal. The more the water moves (larger diffusion coefficient D), and the stronger the sensitizing gradients (larger b), the greater the signal loss. Free water, as in CSF, loses signal rapidly; water restricted by cell membranes, as in acute stroke or dense tumor, loses signal more slowly and appears bright on high-b images.
Key equations
Signal decays mono-exponentially with b-value; D is the diffusion coefficient.
-value for the Stejskal-Tanner pulsed gradient pair: is gyromagnetic ratio, gradient amplitude, pulse duration, pulse separation.
Apparent diffusion coefficient estimated from two images with different b-values.
Derivation
- Spins at position acquire phase proportional to times times times during the first gradient pulse.
- The second pulse applies the opposite phase after time Delta. Stationary spins have net phase zero.
- A spin that has displaced by a random distance has net phase proportional to times its displacement.
- Free diffusion gives Gaussian displacements with variance 2D times the diffusion time.
- Averaging the phase factor over a Gaussian distribution gives an exponential of minus one half the phase variance.
- Collecting the gradient and timing terms gives , with .
Worked example
A voxel has signal 1000 at b=0 and 450 at b=1000 s/mm². Find the ADC, and the expected signal at b=2000 if decay stays mono-exponential.
- Use , so .
- Compute ln(1000/450) = ln(2.222) = 0.799.
- D = 0.799 / 1000 = 7.99e-4 mm²/s.
- At b=2000: S = 1000 × exp(-2000 × 7.99e-4) = 1000 × exp(-1.598).
- exp(-1.598) is about 0.202, so S is about 202.
Answer ADC ≈ 8.0 × 10⁻⁴ mm²/s; predicted signal at b=2000 ≈ 200.
Common mistakes
- Thinking b-value is a property of the tissue; it is set by the scanner sequence.
- Confusing ADC with true diffusion: ADC is an apparent value that includes perfusion, restriction and other effects.
- Forgetting the b=0 reference image when calculating ADC.
- Assuming a higher b-value is always better; signal-to-noise falls and the decay may be non-Gaussian.
- Interpreting T2 shine-through as restricted diffusion without checking the ADC map.