In NMR, nuclei in different electronic environments resonate at slightly different frequencies because electrons shield the nucleus from the applied magnetic field. This difference is reported as the chemical shift in parts per million (ppm), a field-independent scale that reveals molecular structure.
Intuition
Picture a proton sitting in a strong magnetic field . Surrounding electrons are forced by to circulate, and this circulation creates a tiny opposing magnetic field at the nucleus. The nucleus therefore feels slightly less than . The more electron density around it, the greater the shielding and the lower its resonance frequency. A nearby electronegative atom such as oxygen pulls electron density away, so the proton is deshielded, feels a larger field, and resonates at higher frequency (further downfield). Because the shielding effect is proportional to , we divide frequency differences by the spectrometer frequency to get ppm, so spectra from 300 MHz and 600 MHz instruments line up on the same scale.
Key equations
Effective field at the nucleus is reduced by the shielding constant (typically for protons).
Larmor frequency depends on the shielded field; is the gyromagnetic ratio.
Chemical shift in ppm relative to a reference such as TMS.
Derivation
- Start with the Larmor frequency for a bare nucleus: .
- Electrons induce a small opposing field proportional to : .
- The effective field is .
- So .
- For sample and reference: .
- Divide by (the spectrometer frequency) to remove field dependence: approximately.
Worked example
On a 400 MHz spectrometer, a proton resonates 1460 Hz higher than TMS. What is its chemical shift, and what would the offset in Hz be on a 600 MHz instrument?
- .
- The ppm value is field independent, so ppm at 600 MHz.
- Offset at 600 MHz = 3.65 × 600 = 2190 Hz.
Answer 3.65 ppm (typical of a proton on a carbon next to oxygen); 2190 Hz at 600 MHz.
Common mistakes
- Thinking more shielding means higher frequency; it means lower frequency (upfield, smaller ppm).
- Believing ppm values change with field strength; only the Hz separation changes.
- Confusing downfield with a lower ppm; downfield means higher ppm.
- Forgetting that the reference (TMS) is defined as 0 ppm and is highly shielded.