Summary
After an RF pulse tips magnetization into the transverse plane, the spins within a voxel precess at slightly different frequencies and fan out, so the net signal decays faster than true T2 would predict. This decay is characterized by T2*, which combines irreversible T2 relaxation with reversible dephasing from static field inhomogeneity.
Intuition
Picture a voxel as a crowd of small spin groups called isochromats, each a tiny vector in the transverse plane. Right after the pulse they all point the same way and add up to a large signal. But each experiences a slightly different local field because of magnet imperfections, tissue susceptibility differences, and air-tissue boundaries. So each rotates at its own rate. Fast ones race ahead, slow ones lag, and the vectors spread like runners after a starting gun. The vector sum shrinks even though each individual vector keeps its length. Part of this loss is reversible, because a 180-degree pulse can refocus static offsets (a spin echo). The part caused by random molecular interactions fluctuating in time is irreversible and is the true T2. T2* is always shorter than or equal to T2.
Key equations
Observed decay rate is the true T2 rate plus a rate from field inhomogeneity spread (ΔB0 as the field spread across the voxel).
Each isochromat's frequency offset is proportional to its local field offset.
Net transverse magnetization is the vector sum of isochromats with phases accumulating over time, times intrinsic T2 decay.
Free induction decay or gradient-echo signal envelope for a Lorentzian frequency spread.
Derivation
- Start with isochromats all at phase zero immediately after a 90-degree pulse, each with magnitude .
- In the rotating frame at the Larmor frequency, isochromat i has offset Δω_i = γΔB_i and phase φ_i(t) = Δω_i t.
- The net signal is the average of e^{iφ_i}, multiplied by the irreversible factor e^{-t/T2}.
- Assume offsets follow a Lorentzian distribution with half-width Δω_h. The average of e^{iΔω t} is then e^{-Δω_h t}.
- So the envelope is e^{-t(1/T2 + Δω_h)}.
- Define 1/T2* = 1/T2 + Δω_h, with Δω_h = γΔB0 playing the role of the reversible inhomogeneity rate (often written 1/T2').
Worked example
A tissue has T2 = 80 ms. Field inhomogeneity contributes a reversible rate 1/T2' = 15 s^-1. Find T2* and the fraction of signal remaining at TE = 20 ms in a gradient echo.
- 1/T2 = 1/0.080 s = 12.5 s^-1.
- 1/T2* = 12.5 + 15 = 27.5 s^-1.
- T2* = 1/27.5 = 0.0364 s = 36.4 ms.
- Signal fraction = exp(-TE/T2*) = exp(-0.020 × 27.5) = exp(-0.55).
- exp(-0.55) ≈ 0.577.
Answer T2* ≈ 36 ms; about 58% of the initial signal remains at TE = 20 ms (a spin echo would retain exp(-0.25) ≈ 78%).
Common mistakes
- Thinking T2* is a tissue constant like T2; it depends on shimming, voxel size, and field strength.
- Believing dephasing means the individual spins lose magnetization; their magnitudes remain and only phase coherence is lost.
- Assuming a 180-degree pulse recovers all signal; it removes only static offsets, not true T2 loss.
- Assuming T2* can be longer than T2; it is always shorter or equal.
- Adding times instead of rates when combining T2 and T2'.