Content
Summary
A magnetic moment with angular momentum placed in a static magnetic field does not simply align with the field. It precesses around at the Larmor frequency, which is proportional to field strength. This frequency is the resonance frequency that MRI and NMR use to excite and detect signal.
Intuition
Think of a spinning top tilted away from vertical. Gravity tries to topple it, but because it spins, it instead circles around the vertical axis. A nuclear spin behaves similarly: the field exerts a torque on the magnetic moment, but because the moment is tied to angular momentum, the result is a sideways motion, a circling at a fixed angle around . The stronger the field, the faster the circling. The gyromagnetic ratio is a property of the nucleus that sets the proportionality. For hydrogen in a 1.5 T scanner the circling is about 64 million times per second, which is in the radiofrequency range. The angle with stays constant without relaxation, so the longitudinal component is fixed and the transverse component rotates, and a rotating transverse magnetization is what induces a signal in a receiver coil.
Key equations
Torque on the magnetic moment equals the rate of change of its angular momentum, giving precession.
Larmor angular frequency in rad/s.
Larmor frequency in Hz; for H, is about 42.58 MHz/T.
Derivation
- Take along . The moment has angular momentum with .
- Torque equals dJ/dt = mu x B0, so dmu/dt = gamma mu x B0.
- Write components: , , .
- Differentiate again: , a harmonic oscillator.
- Solution: , , with .
- So is constant and the transverse part rotates clockwise viewed from , at .
Worked example
Find the Larmor frequency of hydrogen protons in a 3.0 T scanner, and compare with 1.5 T.
- Use with .
- At 3.0 T: f0 = 42.58 x 3.0 = 127.7 MHz.
- At 1.5 T: f0 = 42.58 x 1.5 = 63.9 MHz.
- Ratio is 2, since is proportional to .
Answer About 127.7 MHz at 3 T, twice the 63.9 MHz at 1.5 T.
Common mistakes
- Thinking spins simply align like compass needles and stop; they precess.
- Confusing angular frequency (rad/s) with frequency (Hz) and forgetting the factor .
- Believing precession frequency depends on the tilt angle; it depends only on and .
- Assuming the Larmor frequency is the same for all nuclei; differs, e.g. is about a quarter of .
- Confusing precession with spin itself; precession is the motion of the moment's axis.