Content
Summary
Absorbance measures how much light a solution absorbs at a given wavelength. The Beer-Lambert law says absorbance is directly proportional to concentration and path length, so a spectrophotometer reading can be used to find an unknown concentration.
Intuition
Imagine shining a flashlight through a glass of tea. A weak tea lets most light through, while a strong tea blocks more. Each dissolved molecule has a small chance of catching a photon, so doubling the number of molecules in the beam doubles the absorbing effect on a logarithmic scale. Absorbance is defined with a logarithm, so each equal layer of solution removes the same fraction of the remaining light, and this turns that multiplicative dimming into a straight-line relationship with concentration. A longer cuvette also puts more molecules in the beam, so path length matters in the same way.
Key equations
Transmittance is the fraction of incident light that passes through the sample.
Absorbance is the negative base-10 log of transmittance; it has no units.
Beer-Lambert law: absorbance equals molar absorptivity times path length times concentration.
Rearranged form to find an unknown concentration.
Derivation
- Consider a thin slice of thickness dx. The light lost in it is proportional to the light arriving and the concentration: dI = -k c I dx.
- Rearrange: dI/I = -k c dx.
- Integrate from 0 to , where intensity goes from to : .
- Convert to base 10: .
- Define the molar absorptivity as and , giving .
Worked example
A dye solution in a 1.00 cm cuvette has absorbance 0.450 at 520 nm. The molar absorptivity at this wavelength is 15,000 L mol^-1 cm^-1. Find the concentration and the percent transmittance.
- Use c = A / (epsilon l).
- c = 0.450 / (15000 × 1.00) = 3.00 × 10^-5 mol/L.
- Transmittance T = 10^(-A) = 10^(-0.450) = 0.355.
- Percent transmittance = 0.355 × 100 = 35.5%.
Answer Concentration = 3.00 × 10^-5 M; transmittance about 35.5%.
Common mistakes
- Confusing absorbance with transmittance: high absorbance means low transmittance.
- Forgetting that absorbance is unitless, or leaving out the path length when it is not 1 cm.
- Using the law at very high concentrations or absorbances above about 1 to 2, where the graph curves and readings are unreliable.
- Not blanking the spectrophotometer with the pure solvent first.
- Measuring at a wavelength different from the one at which epsilon was determined.